Technical data
Command Reference 5
Programmer’s Guide 83
Trace Subsection
Select Trace Display Mode
:TRACe1|2|3|4:MODE WRITe|MAXHold|MIN-
Hold|VIEW|BLANk
:TRACe1|2|3|4:MODE?
Selects the display mode for the selected trace.
WRITe puts the trace in the normal mode, updating the data.
MAXHold displays the highest measured trace value for all the data that
has been measured since the function was turned on.
MINHold displays the lowest measured trace value for all the data that
has been measured since the function was turned on.
VIEW turns on the trace data so that it can be viewed on the display.
BLANK turns off the trace data so that it is not viewed on the display.
Example: :TRAC1:MODE VIEW
Key access: Trace > Clear Write|Max Hold|Min Hold|View|Blank
Trace Max/Min Hold By Count
:TRACe[1]|2|3|4:HOLD:COUNt <value>
:TRACe[1]|2|3|4:HOLD:COUNt?
Use this command to set the Max/Min hold counter.
Example: :TRACe:HOLD:COUNt 20
Key access: Trace > More > Max/Min Number On
Add Trace
:TRACe:MATH:ADD
<destination_trace>,<source_trace1>,<source_trace2>
This command adds the selected trace on the activated trace.
Example: :TRAC:MATH:ADD TRACE2,TRACE1,TRACE3