Technical data
26 Programmer’s Guide
3 Status Registers
The Status Byte Register
The RQS bit is read and reset by a serial poll. The same bit position
(MSS) is read, non-destructively by the *STB? command. If you serial
poll bit 6 it is read as RQS, but if you send *STB it reads bit 6 as MSS.
For more information refer to IEEE 488.2 standards, section 11.
0
1
2
3
4
5
6
7
Status Byte Register
Unused
Unused
Error/Event Queue Summary Bit
Unused
Message Available (MAV)
Standard Event Summary Bit
Unused
Operation Status Summary Bit
&
&
&
&
&
&
&
+
0 1 2 3 4 5 6 7
Service Request
Enable Register