Technical data

212 Programmer’s Guide
5 Command Reference
Set Trigger Level
:TRIGger[:SEQuence]:LEVel <value>
:TRIGger[:SEQuence]:LEVel?
This command defines the trigger level of power sensor. It’s only
available for continuous trigger or single trigger. The trigger source
must be internal, and auto trigger level state must be off (set to normal
mode).
Range: –100 to 0 dBm
Example: :TRIGger:SEQuence:LEVel -20dBm
Key access: Meas > Peak Setup > Trig/Acq > Level
Power Meter Trigger Delay Time
:TRIGger[:SEQuence]:DELay <value>
:TRIGger[:SEQuence]:DELay?
This command defines trigger delay time of power meter measurement.
Range: 0 to 1 s
Example: :TRIGger:SEQuence:DELay 1
Key access: Meas > Peak Setup > Trig/Acq > Delay
Power Meter Trigger Slope Mode
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
This command toggles trigger slope mode between rise and fall.
Example: :TRIGger:SEQuence:SLOPe POSitive
Key access: Meas > Peak Setup > Trig/Acq > More 1 of 2 > Slop