Technical data

Command Reference 5
Programmer’s Guide 201
Power Meter Option Subsystem
The Power Meter Option Subsystem provides you the SCPI command
reference for the Power Meter function. Use the command
:INSTrument POWmeter to access this subsystem.
Option PWM or PWP is required to enable this function. Option PWM
supports Agilent U2000 series USB average power sensor, and Option
PWP supports Agilent U2020 X series peak power sensor.
Power Sensor Zeroing
:CALibration[1]:ZERO:ONCE
This command perform power sensor zeroing operation.
Example: :CAL1:ZERO:ONCE
Key access: Meas > Zeroing > Zeroing
Zeroing Type
:CALibration[1]:ZERO:TYPE EXTernal|INTernal
:CALibration[1]:ZERO:TYPE?
This command toggles the zeroing type between external type and
internal type. U2020 X series peak power sensor only support internal
zeroing type.
Example: :CALibration1:ZERO:TYPE EXTernal
Key access: Meas > Zeroing > Zero Type
Power Meter Sweep State
[:SENSe]:SWEep:STATe RUN|PAUSe
[:SENSe]:SWEep:STATe?
This command toggles power meter sweep state between run and pause.
Example: SWEep:STATe PAUSe
Key access: Meas > Pause/ Run