Technical data

Command Reference 5
Programmer’s Guide 185
ASK Measurement Length
[:SENSe]:DMA:MEASurement:LENGth <value>
[:SENSe]:DMA:MEASurement:LENGth?
Use this command to specify measurement length of ASK demodulation
analysis.
Range: 20 to 1100
Example: :DMA:MEASurement:LENGth 200
Key access: Meas > Meas Length
ASK Trigger Type
:TRIGger:DMA[:SEQuence]:SOURce IMMediate|EXTer-
nal|RFTRigger
:TRIGger:DMA[:SEQuence]:SOURce?
Use this command to specify the trigger type used in current ASK
demodulation analysis.
Range: IMMediate, EXTernal, RFTRigger
Example: :TRIGger:DMA:SOURce EXTernal
Key access: Meas > Trigger
ASK RF Trigger Level
:TRIGger:DMA[:SEQuence]:LEVel:ABSolute <value>
:TRIGger:DMA[:SEQuence]:LEVel:ABSolute?
Use this command to turn on/off scale coupling state of X axis.
Range: –40 to 10 dBm
Example: :TRIGger:DMA:LEVel:ABSolute 5dBm
Key access: Meas > Trigger > RF Trigger > Video Level