Technical data
Command Reference 5
Programmer’s Guide 137
CAT Windows Select
[:SENSe]:CAT:WINDow
NONE|RECTangular|HAMMing|S3BLackman|S4BLackman
[:SENSe]:CAT:WINDow?
This command toggles the measurement window between Rectangular
Window, Hamming Window, 3-Sample Blackman Window and
4-Sample Blackman Window.
Range: NONE, RECTangular, HAMMing, S3BLackman, S4BLackman
Example: :CAT:WINDow RECT
Key access: Meas > More 1 of 2 > Window
CAT Start Frequency
[:SENSe]:CAT:FREQuency:STARt <freq>
[:SENSe]:CAT:FREQuency:STARt?
This command defines start frequency of cable and antenna
measurement.
Range: 5 MHz to 7 GHz
Example: :CAT:FREQuency:STARt 200 MHz
Key access: Freq > Start Frequency
CAT Stop Frequency
[:SENSe]:CAT:FREQuency:STOP <freq>
[:SENSe]:CAT:FREQuency:STOP?
This command defines stop frequency of cable and antenna
measurement.
Range: 5 MHz to 7 GHz
Example: :CAT:FREQuency:STOP 1 GHz
Key access: Freq > Stop Frequency