Technical data
134 Programmer’s Guide
5 Command Reference
Cable and Antenna Test Option Subsystem
The Cable and Antenna Test Option Subsystem provides you the SCPI
command reference for the cable and antenna test function. Use the
command :INSTrument CAT to access this subsystem.
Option CA7 or CAU is required to enable this function.
Below use the shorthand of CAT as Cable and Antenna Test.
CAT Measurement Type
[:SENSe]:CAT REFLection|OPCLoss|DTF
[:SENSe]:CAT?
Toggles the cable and antenna measurement mode between Reflection
Measurement, One Port Cable Loss Measurement and Distance to Fault
Measurement.
Range: REFLection, OPCLoss, DTF
Example: :CAT OPCL
Key access: Meas > Meas > Reflection Measurement / One Port Cable Loss /
Distance To Fault
CAT Calibration Type
[:SENSe]:CAT:CALibrate:TYPE FULL|SELected
[:SENSe]:CAT:CALibrate:TYPE?
This command toggles calibration type between full frequency range
calibration and selected frequency range calibration.
Example: :CAT:CALibrate:TYPE FULL
Key access: Freq > Cal Type