Technical data

Command Reference 5
Programmer’s Guide 99
Marker to Center
:CALCulate:MARKer[1]|2|3|4|5|6[:SET]:CENTer
This command sets the center frequency equal to the specified marker
frequency, which moves the marker to the center of the screen. In delta
marker mode, the center frequency is set to the marker delta value. This
command is not available in zero span. This command is just available
in Spectrum Analyzer and Tracking Generator mode.
Example: :CALCulate:MARKer1:CENTer
Key access: Marker
> Marker To > To Center
Marker to Start Frequency
:CALCulate:MARKer[1]|2|3|4|5|6[:SET]:STARt
Sets the start frequency to the value of the specified marker frequency.
In delta marker mode, the start frequency is set to the marker delta
value. This command is not available in zero span. This command is just
available in Spectrum Analyzer and Tracking Generator mode.
Example: :CALCulate:MARKer:STARt
Key access: Marker
> Marker To > To Start
Marker to Stop Frequency
:CALCulate:MARKer[1]|2|3|4|5|6[:SET]:STOP
Sets the stop frequency to the value of the specified marker frequency.
In delta marker mode, the stop frequency is set to the marker delta
value. This command is not available in zero span. This command is just
available in Spectrum Analyzer and Tracking Generator mode.
Example: :CALCulate:MARKer:STOP
Key access: Marker
> Marker To > To Stop