Key Reference 5
N9340A User’s Guide 87
measuring range for the measurement. When
auto ranging is OFF, the power sensor remains
in the current range and may be manually set
by the user.
• Range
Selects the Lower or Upper power ranges when
the Auto Range is Off.
• Limits
Accesses the Limits submenu.
• Limits
Enables or disables the limit function.
• Upper Limit
Sets the upper test limit value for the current
measurement. The units used are dependent on
the current setting of power unit. When the
measured power is greater than the value
specified, a failure will be reported.
• Lower Limit
Sets the lower test limit value for the current
measurement. The units used are dependent on
the current setting of power unit. When the
measured power is less than the value specified,
a failure will be reported.
• Limit Beep
Enables or disables the limit beep.
• Rel/Offset
Accesses the Rel/Offset submenu.
• Rel
Enables the comparison of a measurement result
to a reference value and sets a reference value
on the current measured data.
• Offset
Sets a value of the display offset on the current
measurement, and enables or disables the
display offset. The display offset is applied to
the measurement signal after any math
calculation.