Making Measurements 4
N9340A User’s Guide 75
predefined upper or lower amplitude limit.
1 Press [Preset] to perform a factory preset.
2 Press [Mode] > {Power Meter} to turn on the power
meter.
3 Press {Meas Setup} > {Limits} to access the limits
menu.
4 Press {Limits} to activate the limits function. Each
time the softkey is pressed, the selected option
changes.
5 Press {Upper Limits} and enter the high limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
6 Press {Lower Limits} and enter the low limit value
using the numeric keypad. Then press {dBm} to
complete the setup.