User`s guide

Table Of Contents
Getting Started 1
N9340A User’s Guide 23
Some Helpful Tips
Performing a time-base calibration
The N9340A provides a manual calibration function
to calibrate the time base.
The analyzer should warm up for approximately 30
minutes before calibration.
Use a BNC cable to connect a 10 MHz reference
signal to the EXT TRIG IN connector of your N9340A,
then press [SYS] > {More (2 of 3)} > {Calibration} > {Time
Base} to initiate a calibration.
Selecting a preset type
The N9340A has two Preset options. Power On Pre-
set, and the Preset hardkey function. Each can be
defined by the user using the System menu. The
analyzer has two types of preset setting for you to
choose from:
Default Restores the analyzer to its
factory- defined setting. The factory default
settings can be found, “Factory Default
Settings" on page 39
User
Restores the analyzer to a user-defined setting.
See “Saving a User- defined Preset" on page 23.
Saving a User-defined Preset
If you frequently use system settings that are not
the factory defaults, refer to the following steps to
create a user- defined system settings that can be
easily recalled at the power up state:
1 Set analyzer parameters by the knob, the arrow
keys or the numeric keypad.
NOTE
The Analyzer will always use the factory-default settings
when powering on.