Specifications

5
The N9340B INM shows spectrogram and
spectrum trace on the same screen.
Extended spectrogrammonitoring (Option INM)
Option INM extends the N9340B spectrogram capability further with the ability
to continuously monitor and save spectrogram data over time not only to the
analyzer’s internal memory or a USB flash drive, but also directly to a PC. With
option INM users can save more than 1,500 continuous frames of data depend-
ing on the size of the USB flash drive or PC memory.
Option INM also supports an Automatic Save function. The N9340B can save
the spectrogram at a user specific time or at a set interval of frame. All the
small files with the same prefix can be combined into one file using N9340 PC
Software, making analysis easier.
The user can choose among the 3 display modes: spectrogram only, spectrum
trace only or the combination of a spectrogram and a spectrum trace in one
screen. Limit lines with pass/fail functionality is also available in this measure-
ment. The N9340B will identify the failed frame of spectrogram data with a
red mark. The Pass/Fail test can also provide alerts. Users may use marker to
find the previous/next frame failed data to quickly find the offending event or
interference. The spectrogram data may be played back for review on either the
analyzer’s display or on a PC utilizing N9340B PC Software.
With the option INM, the Agilent N9340B spectrum analyzer can provide
unattended monitoring of communication systems capturing performance
or intermittent events like interference over extended periods of time – days
rather than hours. The option INMcan provide additional value for the wireless
network communication system managers, hospital administrators, etc. as well
as police and homeland security by recording spectrogram measurement results
over time.