Specifications

22
Options Supplemental information
Spectrogram Monitoring (Option INM)
3 display modes Spectrogram
Spectrum trace
Combination of spectrogram and spectrum trace in one screen
Low frequency performance enhancement and xDSL measurement capability (Option XDM)
XDM channel 9 kHz to 12 MHz
DANL 30 Hz RBW, 3 Hz VBW, 50 Ω termination
on input, 0 dB attenuation, RMS detector,
Trace Average ≥ 40
Preampoff
Reference level ≤ –50 dBm
9 kHz < fc ≤ 100 kHz < –117 dBm, Nominal
10 MHz < f
c ≤ 12 MHz < –132 dBm, Nominal
Preampon
Reference level ≤ –70 dBm
100 kHz < f
c ≤ 1 MHz < –138 dBm, Nominal
1 MHz < f
c ≤ 12 MHz < –140 dBm, Nominal
Low frequency performance enhancement and AM/FM In-Band On-Channel IBOC Measurement (Option IBC)
Frequencyrange
AM channel 430 to 1800 kHz According to IBOC (AM) requirement
FM channel 87.25 to 108.55 MHz According to IBOC (FM) requirement
DANL 30 Hz RBW, 3 Hz VBW, 50 Ω termination
on input, 0 dB attenuation, RMS detector,
Trace average ≥ 40
Preampoff
Reference level ≤ –50 dBm
9 kHz < fc ≤ 100 kHz < –117 dBm, Nominal
10 MHz < f
c ≤ 12 MHz < –132 dBm, Nominal
Preampon
Reference level ≤ –70 dBm
100 kHz < f
c ≤ 1 MHz < –138 dBm, Nominal
1 MHz < f
c ≤ 12 MHz < –140 dBm, Nominal
RF preamplifier (Option PA3)
Frequency range 1 MHz to 3 GHz
Gain 20 dB Nominal
Tracking generator (Option TG3)
Frequency range 5 MHz to 3 GHz
Output level 0 to –25 dBm 1 dB steps
Output flatness ± 3 dB Referenced to 50 MHz, 0 dBm
VSWR < 2.0 : 1 Nominal
Connector and impedance Type-N female, 50 Ω