User`s guide

7 Error Messages
144 N9340A User’s Guide
162 Cannot open fast
sweep in this
case
Indicates current settings do not
allow you to open fast sweep.
For example, currently in FFT,
zero span, or any measurement
in power suite does not allow
fast sweep.
159 Error 0x11, DSP
will restart
Indicates DDC failed. If analyzer
froze, try to restart the analyzer.
158 Error 0x10, DSP
will restart
Indicates an internal error
occured, wait the analyzer
to reboot. If analyzer froze, try
to restart the analyzer.
157 Error 0xF Indicates an error in internal
communication.
156 Incorrect
alignment file
on flash
Indicates an invalid alignment
file.
153 RF EEPROM
operate fail
Indicates an error when writing
alignment file.
152 Alignment file
oversize
Indicates the alignment file has
error or invalid data.
151 DSP boot fail Indicates the analyzer
cannot process
measurement,wait the
analyzer to reboot. If
analyzer froze, try to restart the
analyzer.
150 Mixer overload Indicates the first mixer in
danger. Either increase input
attenuation or decrease the input
signal level.
149 Mixer saturate Indicates you need to either
increase the input attenuation or
decrease the input signal level.
148 Error 0xE Indicates the analyzer reads
unreasonable parameter value. If
analyzer froze, try to restart the
analyzer.
Error Code Error Message Description