User`s guide

6 SCPI Command Reference
112 N9340A User’s Guide
DISPlay Subsystem
The DISPlay subsystem controls the selection and
presentation of textual, graphical, and trace
information. Within a display, information may be
separated into individual windows.
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
This command turns the display on or off. Having
the display turned off may increase repetitive
measurement rate.
Trace Y-Axis Scaling
:DISPlay:WINDow:TRACe:Y[:SCALe]
:PDIVision 1|2|5|10
:DISPlay:WINDow:TRACe:Y[:SCALe]
:PDIVision?
This command sets the per- division display scaling
for the y- axis when y- axis units are set to
amplitude units.
*RST State: On
*RST Value: 10 dB
Range: 1 dB, 2 dB, 5 dB or 10 dB
Key Access: [AMPTD] > {Scale/DIV}