User`s guide
3 System and File
24 N9340B User’s Guide
System Setting
This section provides you the basic system
configuration which is frequently used before or
after the measurement operation. Please refer to
the description below to get a better user
experience and ease your test and measurement.
Power On and Preset Settings
Selecting a preset type
Press [SYS] > {Setting} > {PwrOn/Preset} > {Preset Type}
to choose the preset types. The analyzer has three
types of preset setting for you to choose from:
DFT Restores the analyzer to its factory- defined
setting. The factory default settings can be
found, “Factory Default Settings" on page 25.
User Restores the analyzer to a user- defined set-
ting. Refer to the descriptions as below.
Last Restores the analyzer to the last time setting.
Saving a User-defined Preset
If you frequently use system settings that are not
the factory defaults, refer to the following steps to
create a user- defined system settings that can be
easily recalled at the power up state:
1 Set analyzer parameters by the knob, the arrow
keys or the numeric keypad.
2 Press [SYS] > {Setting} > {PwrOn/Preset} > {Save
User} to save the current parameters as the user
preset setting.
3 Press [SYS] > {Setting} > {PwrOn/Preset} > {Preset
Typ e U se r} to set the preset mode to user defined
system setting.
4 Press [Preset], the instrument will be set to the
state you previously saved.