User`s guide

5Key Reference
110 N9340B User’s Guide
Palette
Key access: [MEAS]> {Spectrogram} > {Palette}
Toggles the spectrogram color mode between full
color and gray scale.
Disp Mode
Key access: [MEAS]> {Spectrogram} > {display}
Allows you to set the spectrogram display mode as
trace, spectrogram or both trace and spectrogram.
Save&Load
Key access: [MEAS]> {Spectrogram} > {Save&Load}
This softkey is only available with option INM
activated.
Start/Stop Save
Key access: [MEAS]> {Spectrogram} > {Save&Load} >
{Start/Stop Save}
This softkey allow you to start or stop the
automatic save function. Once the spectrogram
measurement frames reached the file size set in
the {File Setup}, a file (*.SPE or *.CSV) will be
automatically saved, and the spectrogram
measurement will restart.
Timed Setting
Key access: [MEAS]> {Spectrogram} > {Save&Load} >
{Time Setting}
This submenu allows you to configure automatic
time save function. You can set start date, time,
stop date, time in this menu. Press {Timed Save On},
the N9340B will start the auto- saving function
from the Start Time.
NOTE
The save process takes a few seconds to read from cache.
During this period, the measurement is paused.