User`s guide

Key Reference 5
N9340B User’s Guide 107
Specifies the resolution bandwidth used in
measuring the offset pair. When set to Auto, the
Res BW from the default tables for
“Offset/Limits" on page 104 are used. When set to
Man, the allowed range is 1 kHz to 1MHz.
Abs StartAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enables you to enter an absolute level limit at Start
Freq for the currently selected offset ranging from
–200.00 to +50.00 dBm with 0.01 dB resolution.
Abs StopAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enters an absolute level limit at Stop Freq for the
currently selected offset ranging from –200.00 to
+50.00 dBm with 0.01 dB resolution, and to toggle
this function between Couple and Man. If set to
Couple, this is coupled to Abs Start to make a flat
limit line. If set to Man, Abs Start and Abs Stop you
can enter different values to make a sloped limit
line.
Rel StartAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enters a relative level limit at Start Freq ranging
from –150.00 to +50.00 dBc with 0.01 dB
resolution.
Rel StopAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enters a relative level limit at Stop Freq ranging
from –150.00 to +50.00 dBc with 0.01 dB
resolution, and to toggle this function between
Couple and Man. If set to Couple, this is coupled to
Rel Start to make a flat limit line. If set to Man, Rel
Start and Rel Stop you can enter different values to
make a sloped limit line.