User`s guide
Key Reference 5
N9340B User’s Guide 99
Marker Trace
Key access: [MARKER] > {More} > {Marker Trace}
Activates a marker on the trace if there are no
markers turned on. If a marker is currently active,
press Marker Trace until 1, 2, 3, or 4 is underlined.
The active marker will be moved to the selected
trace.
Selecting Auto will move the marker to the
automatically selected trace. The selection order is
to look for the lowest numbered trace in the
following order of trace modes:
If there is no trace in any of these modes, it
selects trace 1 as default.
All Off
Key access: [MARKER] > {More} > {All Off}
Turns off all of the markers, including markers
used for signal track and demodulation. The
marker annotation is also removed.
Marker Table
Key access: [MARKER] > {More} > {Marker Table}
Compresses the graticule and displays the marker
number, trace number, marker type, X axis value,
and the amplitude in the table.
NOTE
Frequency counter is properly effective only on CW signals or
discrete peaks. For a valid reading, the marker must be greater
than 25 dB above the noise. Freq Count function is not
supported under Marker Delta condition.
NOTE
Noise mode operates only on noise peaks and not supported
under Marker Delta condition.
Clear write Max hold
Min. hold View