Peak Search
Place a marker on the highest peak based on the
setting of the Search Criteria as follows:
•Peak
Key access: [MARKER] > {Peak Search}
Place a marker on the highest peak.
•Next Left Peak
Key access: [MARKER] > {Peak Search} > {Next Left PK}
Moves the marker to the next peak to the left of
the current marker. The signal peak must exceed
the peak threshold value by the peak excursion
value. If there is no peak to the right, the marker
will not move and the No Peak Found error message
will appear on the display.
•Next Right Peak
Moves the marker to the next peak to the right of
the current marker. The signal peak must exceed
the peak threshold value by the peak excursion
value. If there is no peak to the right, the marker
will not move and the No Peak Found error message
will appear on the display.
Key access: [MARKER] > {Peak Search} > {Next Right
PK}
•6 Peaks Search
In non- zero span, pressing 6 Peaks Search turns all
six markers on and puts them on the six highest
peak in the current frequency span.
In zero span, the function is not available.
Key access: [MARKER] > {Peak Search} > {Auto Search}