User`s guide

78 N9320A Spectrum Analyzer
4 Key Reference
Det/Display
Press this hardkey to control detector and display functions.
Detector
Auto Man
Selects a specific detector, or allows to pick the appropriate
detector (through Auto) for a particular measurement.
When discussing detectors, it is important to understand
what is a trace “bucket”. For every trace point displayed,
there is a finite time during which the data for that point is
collected. The analyzer look at all of the data collected
during that time and present a single point of trace data
based on the detector mode. We call the interval during
which the data for that trace point is being collected, the
bucket”. Thus a trace is more than a series of single points.
It is actually a series of trace “buckets”. The data may be
sampled many times within each bucket.
Figure 3 Detector (auto) Selection Flowchart Rules
Phase
Noise On
Res BW
>300Hz?
Tra ce
Average On?
Sample
Average
Yes
Yes
No
No
Avg Type
Power?
Sample
Average
Yes
Yes
No
No
Any Trace in
MinHold
Peak
Sample
Yes
Yes
No
Any Trace in
Max Hold?
No
Peak
Negative Peak
Yes
Any Trace in
Min Hold?
No