User`s guide

70 N9320A Spectrum Analyzer
4 Key Reference
Amplitude
Press activates the reference level function and
accesses the amplitude softkeys.
Ref Level Sets the reference level. This function is activated when
is pressed. The reference level is the amplitude power
or voltage represented by the top line of the graticule on the
display. Setting the reference level changes the absolute
amplitude level (in the selected amplitude units) of the top
graticule line.
Key access:
Attenuation
Auto Man
Toggles the input attenuation between Auto and Man and set
the attenuation level in 5 dB increments when Man is
selected. The analyzer input attenuator, which is normally
coupled to the reference level control, reduces the power
level of the analyzer input signal at the input mixer.
Selecting Auto re- couples the attenuator. Change the
attenuation by using the arrow keys, the knob, or the
numeric keypad.
Key access:
Scale/Div Sets the logarithmic units per vertical graticule division on
the display. The Scale/Div function is available only when the
Scale Type key is set to Log.
Key access:
Scale Type
Log Lin
Scales the vertical graticule divisions in logarithmic units
when Log is selected. The valid logarithmic units are 1, 2, 5
and 10 dB per division. When Lin is underlined, the vertical
divisions are linearly scaled with default amplitude units of
volts. The top of the graticule represents the reference- level
value; the bottom of the graticule represents zero volts.
(Each division is one- tenth of the reference level.)
Key access:
Amplitude
Amplitude
Amplitude
Amplitude
Amplitude
Amplitude