If the signal peak is not present on the display, span out to
20 MHz, turn signal tracking on, span back to 2 MHz and
turn signal tracking off:
Press > Span > 20 > MHz
Press
Press > Signal Track (On)
Press > 2 > MHz
Press > Signal Track (Off)
Figure 5 Unresolved Signals of Equal Amplitude
4 Change the resolution bandwidth (RBW) to 100 kHz so that
the RBW setting is less than or equal to the frequency
separation of the two signals:
Press > 100 > kHz
Notice that the peak of the signal has become flattened
indicating that two signals are present.
5 Decrease the video bandwidth to 3 kHz:
Press > Video BW > 3 > kHz
Two signals are now visible as shown in Figure 6. Use the
front- panel knob or arrow keys to further reduce the
resolution bandwidth and better resolve the signals.