User`s guide

22 N9320A Spectrum Analyzer
2 Getting Started
Some helpful Tips
Please refer to the following descriptions for your further
use, including alignment, external reference, firmware update
and option activation.
Running Internal Alignments
To meet specifications, the analyzer must be manually
aligned.
Manually Performing an Alignment
1 Connect a BNC cable with N- BNC adapter between the CAL
OUT and RF IN connector.
2 After warm- up, press > Alignment > Align > All.
Please refer to “Alignment” on page 109 for details.
Enable the Options
You are required to enter the option license key to enable
the option. Contact your nearest Agilent Office for
purchasing a license. Please refer to the procedures below to
activate the options you have purchased. (option TG3 and
option PA3)
1 Press > More > More > Licensing > Option
2 Enter the option number which you want to enable, Press
[Enter] to confirm your input.
3 Press > More > More > Licensing > License key
4 Enter
and terminate the license input. The option will be
enabled immediately.
NOTE
When the alignment runs, you will hear clicking as the attenuator
settings changing. This is not an indication of trouble.