User`s guide
124 N9320A Spectrum Analyzer
5 One-Button Measurements
weights new data more than old data, which facilitates
tracking of slow- changing signals. The average is displayed at
the end of each sweep.
When you select Repeat, after reaching the average count, all
previous data is cleared and the average count is back to 1.
Key access:
> ACP
Chan Integ BW Specifies the range of integration used in calculating the
power in the main channel. Use the knob and the arrow
keys to set the bandwidth.
Key access:
> ACP
Offset/Limits Accesses softkeys that allow you to configure the offsets for
the ACP measurement.
Key access:
> ACP
Meas
Meas
Offset
A B C D E F
Selects the offset the softkeys will affect. Press
Offset until the letter of the desired offset is
underlined.
Key access:
> ACP > Offset/Limits
Offset Freq
On Off
Sets the frequency difference from the center
of the main channel to the center of the offset
for a maximum of 6 offsets (labeled A- F). It
also allows you to turn On/Off the offsets that
you want to measure.
Key access:
> ACP > Offset/Limits
Ref BW Sets the reference bandwidth (integration
bandwidth) for each offset.
Key access:
> ACP > Offset/Limits
Pos Offset
Limit
Sets the upper limit for the upper segment of
the specified offset pair.
Key access:
> ACP > Offset/Limits
Neg Offset
Limit
Sets the upper limit for the lower segment of
the specified offset pair.
Key access:
> ACP > Offset/Limits
Meas
Meas
Meas
Meas
Meas
Meas