User`s guide

One-Button Measurements 5
N9320A Spectrum Analyzer 119
Avg Number
On Off
Specifies the number of measurement averages used when
calculating the measurement result with Avg Number (On). The
range of the number is from 1 to 1000. Use the knob, the
arrow keys or the numeric keypad to set the number. The
average will be displayed at the end of each sweep. Selecting
Off disables measurement averaging.
Key access:
> Channel Power
Avg Mode
Exp Repeat
Toggles the average mode between Exp or Repeat. This
determines the averaging action after the specified number
of measurements (average count) is reached.
When you select Exp, each successive data acquisition after
the average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging
weights new data more than old data, which facilitates
tracking of slow- changing signals. The average will be
displayed at the end of each sweep.
If you select Repeat, after reaching the average count, all
previous data is cleared and the average count is back to 1.
Key access:
> Channel Power
Integ BW Specifies the range of integration used in calculating the
power in a channel. For example, set the main (center)
channel bandwidth. Note that the white frame indicates the
integration bandwidth. Be sure the span is set between 1
and 10 times the integration bandwidth. Change integration
bandwidth by using the knob and the arrow keys.
Key access:
> Channel Power
Chan Pwr Span Sets the analyzer span for the channel power measurement
by using the knob and the arrow keys. Be sure the span is
set between 1 and 10 times the integration bandwidth.
Key access:
> Channel Power
Meas
Meas
Meas
Meas