User`s guide

Key Reference 4
N9320A Spectrum Analyzer 107
Preset/System
Accesses the Preset and System softkeys. When the analyzer
is in the remote mode, pressing returns the analyzer
to the local mode and enables front- panel control.
Preset Provides a known state of the instrument for making
measurements. The analyzer allows you to select whether a
user preset, or factory preset is performed.
If Preset Type is set to Factory, pressing > Preset performs
a factory preset, which performs the following:
Resets the analyzer to Spectrum Analyzer (SA) mode
Brings up the Frequency menu
Sets certain conditions to their default values
Clears both the input and output buffers and clears all
trace data
Sets the amplitude values of trace 2, 3 and 4 to the
bottom of the screen
Turn amplitude- correction factors off, but remain in
analyzer memory
Turn limit line testing off, but the limit line tables remain
in analyzer memory
See Table Factory Preset States as follows for the conditions
established by performing a factory preset.
When Preset Type is set to User, pressing > Preset
displays the softkeys User Preset and Factory Preset, you then
select the specific preset as you need.