User`s guide

Key Reference 4
N9320A Spectrum Analyzer 103
Peak Search
Places a marker on the highest peak based on the settings of
the “Search Criteria” on page 104. Refer to “Peak Search
Type on page 105, for more information on the effect of
settings.
Peak Search Performs peak search as described above.
Key access:
Next Peak Places the marker on the next highest peak. The signal peak
must exceed the peak threshold value by the peak excursion
value. If there is no peak, the marker will not move.
Key access:
Next Pk Right Moves the marker to the next peak to the right of the
current marker. The signal peak must exceed the peak
threshold value by the peak excursion value. If there is no
peak to the right, the marker will not move and the “ No
Peak Found error message will appear on the display.
Key access:
Next Pk Left Moves the marker to the next peak to the left of the current
marker. The signal peak must exceed the peak threshold
value by the peak excursion value. If there is no peak to the
left, the marker will not move and the “No Peak Found error
message will appear on the display.
Key access:
NOTE
All peak search functions ignore the LO feed through. The process for
determining if the Peak is the LO feed through takes into account Start
Frequency, Span, Resolution Bandwidth, Resolution Bandwidth shape
factor, and Frequency Accuracy. If there is no point of inflection within the
LO feed through range, the point on the trace with maximum amplitude,
that is higher in frequency than the LO feed through, will be selected.
Peak
Search
Peak
Search
Peak
Search
Peak
Search