User`s guide

102 N9320A Spectrum Analyzer
4 Key Reference
Ref Trace
View Blank
Allows you to view or blank the reference
trace on the display.
Key access:
> Tracking Generator > More
> Normalize
Tracking
Peak
Activates a routine that automatically adjusts fine
frequency tracking to obtain the peak response of the
tracking generator on the spectrum analyzer display.
Tracking Peak is performed in the current resolution
bandwidth.
Key access: > Tracking Generator>More
Man Track
adj
Allows you to adjust the frequency of the tracking
generator oscillator manually using the step keys, knob,
or numeric keypad. The tracking adjust is tuned to
maximize the amplitude of the trace.
Key access:
> Tracking Generator>More
MODE
MODE
MODE
NOTE
When in the Tracking Generator mode, resolution bandwidths < 1 kHz
are not available.