User`s guide
Key Reference 4
N9320A Spectrum Analyzer 97
Marker->
Press accesses the following marker function softkeys:
Mkr CF Sets the frequency of the marker to the center frequency. In
delta mode, Mkr CF sets the center frequency to the
marker delta value. Mkr CF is not available in zero span.
Key access:
Mkr CF Step Changes the center- frequency step size to match the value of
the active marker. Press > CF Step to view the step
size. The step size is the frequency difference between the
markers when marker delta is active. Use this function to
step from one signal harmonic to another. Mkr CF Step is
not available in zero span.
Key access:
Mkr Start Changes the start frequency so that it is equal to the
frequency of the active marker. In delta mode, Mkr Start
sets the start frequency to the marker delta value.
Mkr Start is not available in zero span.
Key access:
Mkr Stop Changes the stop frequency so that it is equal to the
frequency of the active marker. In Delta mode, Mkr Stop
sets the stop frequency to the marker delta value. Mkr Stop
is not available in zero span.
Key access:
Mkr Ref Lel Changes the reference level to the active marker value,
moving the marked point to the reference level (top line of
the graticule). In delta mode, Mkr Ref Lel sets the reference
level to the amplitude difference between the markers. (For
example, if the difference is –15 dB, the reference level –15
dBm.)
Key access:
Marker
Marker
Frequency
Marker
Marker
Marker
Marker