Technical data
92 N9320A Programmer’s Guide
5 Command Reference
SEM measurement
:READ:SEMask?
Returns 20 scalar results, each five in a array, in the
following order:
1. The emission mask offset of lower value displayed. Used
for store the lower statistical result for power(or PSD) offset.
2. The emission mask offset of lower frequency. Used for
store the lower statistical result for frequency offset.
3. The emission mask offset of upper value displayed. Used
for store the upper statistical result for power/PSD offset.
4. The emission mask offset of upper frequency. Used for
store the upper statistical result for frequency offset.
Key Access > SEM
TOI measurement
:READ:TOIntercept?
Returns 18 scalar results, in the following order.
1. The lower base frequency value in Hz.
2. The lower base power value in dBm.
3. The lower base power value in dBc.
4. The upper base frequency value in Hz.
5. The upper base power value in dBm.
6. The upper base power value in dBc.
7. The worst case Output Intercept Frequency value in Hz.
8. The worst case Output Intercept Power value in dBm.
9. The worst case Output Intercept Power value in dBc.
10. The worst case Output Intermod Point in Hz
11. The lower Output Intercept Frequency value in Hz.
12. The lower Output Intercept Power value in dBm.
13. The lower Output Intercept Power value in dBc
14. The lower Output Intermod Point in Hz
15. The upper Output Intercept Frequency value in Hz.
Meas