Load a Limit Line from Memory to the Instrument
:MMEMory:LOAD:LIMit LLINE1|LLINE2,<file_name>
Loads a limit line, from the specified file in mass storage to
the instrument. Loading a time limit line deletes any
frequency limit lines. Similarly, loading a frequency limit line
deletes any time limit lines.
Example: :MMEM:LOAD:LIM LLINE2,mylimit.lim
Remarks: There is no SCPI short form for parameters
LLINE1|LLINE2.
Key access: > Load > Type > Limits
Load an Instrument State from a File
:MMEMory:LOAD:STATe 1,<file_name>
The contents of the state file are loaded into the current
instrument state.
Example: :MMEM:LOAD:STAT 1,mystate.sta
Remarks: You must be in Spectrum Analysis mode to use this command.
See also commands :MMEMory:LOAD:STATe and
:MMEMory:STORe:STATe
If the revision of the state being loaded is newer than the
revision of the instrument, no state is recalled and an error
is reported.
If the revision of the state being loaded is equal to the
revision of the instrument, all regions of the state will be
loaded.
If the revision of the state being loaded is older than the
revision of the instrument, the instrument will only load the
older regions of the state.
Key access: > Load > Type > State