Technical data
82 N9320A Programmer’s Guide
5 Command Reference
for store the lower statistical result for power(or PSD) offset.
2. The emission mask offset of lower frequency. Used for
store the lower statistical result for frequency offset.
3. The emission mask offset of upper value displayed. Used
for store the upper statistical result for power/PSD offset.
4. The emission mask offset of upper frequency. Used for
store the upper statistical result for frequency offset.
Key Access > SEM
TOI measurement
:FETCh:TOIntercept?
Returns 18 scalar results, in the following order.
1. The lower base frequency value in Hz.
2. The lower base power value in dBm.
3. The lower base power value in dBc.
4. The upper base frequency value in Hz.
5. The upper base power value in dBm.
6. The upper base power value in dBc.
7. The worst case Output Intercept Frequency value in Hz.
8. The worst case Output Intercept Power value in dBm.
9. The worst case Output Intercept Power value in dBc.
10. The worst case Output Intermod Point in Hz
11. The lower Output Intercept Frequency value in Hz.
12. The lower Output Intercept Power value in dBm.
13. The lower Output Intercept Power value in dBc
14. The lower Output Intermod Point in Hz
15. The upper Output Intercept Frequency value in Hz.
16. The upper Output Intercept Power value in dBm.
17. The upper Output Intercept Power value in dBc.
18. The upper Output Intermod Point in Hz.
:FETCh:TOIntercept:IP3?
This command returns the worst case Output Intercept
Power value in dBm.
Key Access > TOI
Meas
Meas