Technical data
76 N9320A Programmer’s Guide
5 Command Reference
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
:DISPlay:ENABle?
Turns the display on or off. Having the display turned off
may increase repetitive measurement rate.
Pressing (local) key will turn display enable back on.
Trace X-Axis Scale Offset
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet <freq>
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet?
Specifies the frequency offset for all frequency readouts such
as center frequency, except that it does not affect marker
count.
*RST: 0 Hz
Range: –500 Hz to 500 Hz
Default Unit: Hz
Remarks: Frequency offset is not available when frequency scale type
is Log ([:SENSe]:SWEep:SPACing LINear|LOGarithmic).
Key Access: > Freq Offset
Display Line Amplitude
:DISPlay:WINDow:TRACe:Y:DLINe <ampl>
:DISPlay:WINDow:TRACe:Y:DLINe?
Defines the level of the display line, in the active amplitude
units if no units are specified.
*RST: 2.5 divisions below the reference level
Range: 10 display divisions below the reference level to the
reference level
Default Unit: Current active units
Key access: > Display Line On Off
Preset/
System
Frequency
Det/
Display