Technical data

Command Reference 5
N9320A Programmer’s Guide 71
CALibration Subsystem
These commands control the self- alignment processes.
Align All Instrument Assemblies
:CALibration[:ALL]
:CALibration[:ALL]?
Performs an alignment of all the assemblies within the
instrument, except for the tracking generator.
Before executing this command, connect a cable between
front panel CAL OUT and RF IN connector.
The query performs a full alignment and returns a number
indicating the success of the alignment. A “1” is returned if
the alignment is successful.
Key access: > Alignments > Align > All
Coarse Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:COARse <set-
ting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 20 MHz reference
oscillator timebase of the analyzer.
Range: Integer, 0 to 255
Key access: > Alignments > Time Base
Preset/
System
NOTE
:CALibration:ALL is required after COARse is set.
Preset/
System