Technical data
60 N9320A Programmer’s Guide
5 Command Reference
Frequency Counter Marker Automatic Resolution
:CALCulate:MARKer:FCOunt:RESolution:AUTO
OFF|ON|0|1
:CALCulate:MARKer:FCOunt:RESolution:AUTO?
Sets the resolution of the marker frequency counter so it is
automatically coupled to the frequency span, generating the
fastest accurate count.
*RST: On
Key access:
> Function > Freq Counter > Resolution Auto Man
Frequency Counter Marker
:CALCulate:MARKer[n]:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer[n]:FCOunt[:STATe]?
Turns on or off the marker frequency counter. To query the
frequency counter, use
:CALCulate:MARKer[1]:FCOunt:X? If the specified
marker number is not the active marker, it becomes the
active marker. If the specified marker number is not on, it is
turned on and becomes the active marker. A 1 is returned
only if marker count is on and the selected number is the
active marker.
*RST: Off
Remarks: If a frequency count x value is generated when the
frequency count state is off, then 0 is returned.
Key access:
> Function > Freq Counter > Freq Counter
Frequency Counter Marker Query
:CALCulate:MARKer[n]:FCOunt:X?
Queries the marker frequency counter.
Remarks: If a frequency count x value is generated when the
frequency count state is off, then 0 is returned.
Marker
Marker