Technical data

Command Reference 5
N9320A Programmer’s Guide 53
CALCulate Subsystem
This subsystem is used to perform post- acquisition data
processing. In effect, the collection of new data triggers the
CALCulate subsystem. In this instrument, the primary
functions in this subsystem are markers and limits.
CALCulate subsystem commands used for measurements in
the MEAS menus are located in “SENSe Subsystem” on
page 94.
NdBpoints
:CALCulate:BWIDth|BANDwidth:NDB <rel_ampl>
:CALCulate:BWIDth|BANDwidth:NDB?
Selects the power level, below the peak of the signal, at
which the signal bandwidth will be measured by the
markers.
:CALCulate:BWIDth|BANDwidth[:STATe] must be ON.
*RST: –3 dB
Range: –80 dB to –1 dB
Remarks: Refer to :CALCulate:BWIDth|BANDwidth[:STATe] for an
explanation of this marker function.
Key Access: > More > N dB Points
NdBresults
:CALCulate:BWIDth|BANDwidth:RESult?
Returns the measured bandwidth at the power level defined
by :CALCulate:BWIDth:NDB?. 0 is returned if
:CALCulate:BWIDth|BANDwidth[:STATe] is off, or when
a result is not available. Refer to
CALCulate:BWIDth|BANDwidth[:STATe] for an
explanation of this marker function.
Default Unit: Hz
Key access: > More > N dB Points
Peak
Search
Peak
Search