Video Trigger Level Amplitude
:TRIGger[:SEQuence]:VIDeo:LEVel <ampl>
:TRIGger[:SEQuence]:VIDeo:LEVel?
Specifies the level at which a video trigger will occur.
*RST: 2.5 divisions below reference level
Range: 10 display divisions below reference level to reference level
Default Unit: current amplitude units
Remarks: Video is adjusted using this command, but must also be
selected using the command
:TRIGger[:SEQuence]:SOURce VIDeo.
Key access: > More > Video
Video Trigger Level Amplitude
:TRIGger[:SEQuence]:VIDeo:LEVel <ampl>
:TRIGger[:SEQuence]:VIDeo:LEVel?
Specifies the level at which a video trigger will occur.
*RST: 2.5 divisions below reference level
Range: 10 display divisions below reference level to reference level
Default Unit: current amplitude units
Remarks: Video is adjusted using this command, but must also be
selected using the command
:TRIGger[:SEQuence]:SOURce VIDeo. When in FM
Demod and Demod View is on, the Video Trigger level is
adjusted/queried using the command
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency