Technical data

132 N9320A Programmer’s Guide
5 Command Reference
Persistent State Reset
:SYSTem:PRESet:PERSistent
Sets the persistent state values to their factory defaults. The
term persistent means that the command retains the setting
previously selected, even through a power cycle. Examples of
persistent functions are: power- on type, and preset type.
Preset Type
:SYSTem:PRESet:TYPE FACTory|USER|LAST
Selects the preset state to be either factory- defined or
user- defined preset conditions.
*RST: The factory default is Factory. This parameter is persistent,
which means that it retains the setting previously selected,
even through a power cycle.
Remarks: :SYSTem:PRESet:USER:SAVE defines the user preset.
Key access: > Pwr On/Preset > Preset Type
Save User Preset
:SYSTem:PRESet[:USER]:SAVE
Saves the current instrument conditions as the user preset
condition.
Key access: > Power On/Preset > Save Type Preset
Set Time
:SYSTem:TIME <hour>,<minute>,<second>
:SYSTem:TIME?
Sets the time of the real- time clock of the instrument.
Hour must be an integer 0 to 23.
Minute must be an integer 0 to 59.
Second must be an integer 0 to 59.
Key access: > More > Time/Date > Set Time
Preset/
System
Preset/
System
Preset/
System