Technical data
Command Reference 5
N9320A Programmer’s Guide 125
SOURce Subsystem
The SOURce subsystem controls the signal characteristics of
the tracking generator. Refer also to the “OUTPut
Subsystem” on page 89 which contains a command that
controls the tracking generator output.
Sets the Output Power Offset Correction
:SOURce:CORRection:OFFSet <rel_ampl>
:SOURce:CORRection:OFFSet?
Specifies an offset for the displayed output power level. An
offset power level can be added to the displayed level to
compensate for system losses (for example, cable loss) or
gains (for example, preamplifier gain.) This offset does not
change the power out of the source, it only changes the
display so that it reads out the actual power delivered to the
device under test.
*RST: 0 dB
Range: –327.6 dB to 327.6 dB
Default Unit: Currently selected source power units
Key Access:
> Tracking Generator > Amptd Offset
Source Attenuation
:SOURce:POWer:ATTenuation <ampl>
:SOURce:POWer:ATTenuation?
Attenuates the source output level. Specifically setting
:SOURce:POWer:ATTenuation <ampl> sets the mode to
manual (:SOURce:POWer:ATTenuation:AUTO OFF).
*RST: 18 dB
Range: 0 dB to 60 dB in 10 dB steps
Default Unit: dB
Key Access:
> Tracking Generator > Attenuation Auto Man
MODE
MODE