Technical data

124 N9320A Programmer’s Guide
5 Command Reference
[:SENSe]:SWEep Subsection
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the
display.
*RST: 100.2 ms
Default Unit: seconds
Remarks: A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X- axis represents time rather than
frequency.
Key access: > Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current
settings.
*RST: On
Key access: > Sweep Time Auto Man
Sweep/
Tri g
Sweep/
Tri g