Technical data

Command Reference 5
N9320A Programmer’s Guide 119
Selects auto or manual mode for video bandwidth to
resolution bandwidth ratio.
Refer to User’s Guide for a flowchart that illustrates VBW
and RBW Ratio auto rules.
*RST: Auto
Key access: > VBW/RBW Auto Man
[:SENSe]:DETector Subsection
Automatic Detection Type Selected
[:SENSe]:DETector:AUTO OFF|ON|0|1
[:SENSe]:DETector:AUTO?
Switches automatically to the optimum detection type for
typical measurements using the current instrument settings.
The detector type is average if any of these are on:
Trace averaging when the Average Type is Power (RMS).
The detector type is sample if any of the following
conditions are true:
Trace averaging is on with average type of video
Both max and min hold trace modes are on
Resolution bandwidth is less than 1 kHz, and trace
averaging is on
The detector type is negative peak if any trace is in min
hold and no traces are in max hold.
The detector type is peak if the above conditions are off.
Manually changing the detector function turns Auto off.
Refer to Figure 20 on User’s Guide, which shows a decision
tree of how detection type is determined.
*RST: On
Key access: >
Detector
BW/
Avg
Det/
Display