Technical data
Command Reference 5
N9320A Programmer’s Guide 115
Set the Sweep Time
[:SENSe]:SEMask:SWEeptime <number>
[:SENSe]:SEMask:SWEeptime?
[:SENSe]:SEMask:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:SWEeptime:AUTO?
Specifies the sweep time used in measuring the power in the
main channel.
*RST: Auto
Range: 8.37 ms through 4 ks
Example: SEM:SWE 4 s
Key access:
> More> Spectrum Emission Mask > Ref Channel > Sweep Time
Set the Meas Type
[:SENSE]:SEMask:TYPE TPRef | PSDRef
[:SENSE]:SEMask:TYPE?
Displays a menu where you can select a measurement
reference type, Tota l Pwr R e f or PSD Ref.
*RST: Total Pwr Ref
Example: SEM:TYPE TPRef or SEM:TYPE PSDRef
Key access:
> More> Spectrum Emission Mask > Meas Type
[:SENSe]:AVERage Subsection
Set the Average Count
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Specifies the number of measurements that are combined.
*RST: 1
Range: 1 to 1000
Key access: > Average On Off
Meas
Meas
BW/
Avg