Technical data

Command Reference 5
N9320A Programmer’s Guide 111
Set the Stop Freq
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STOP
<freq>,...[<freq>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STOP?
Specifies the inner limit (frequency closest to the carrier) for
both segments of the specified offset pair. When Start Freq
(Off) is selected, the offset pair is not measured.
Offsets that are turned off, for the currently selected offset
will return a - 999.0 when queried.
*RST: Dependent upon Offset selected, please refer to the User’s
Guide for more details.
Range: 0 Hz to the Stop Freq (for that offset) minus 10 Hz
Example: SEM:OFFS:LIST:FREQ:STOP 2 MHz
Key access:
> More > Spectrum Emission Mask > Offset/Limits > Stop Freq
Set the Sweep Time
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime <time> ...
[<time>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime?
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:SWEeptime:AUTO?
Specifies the sweep time for the currently offset, and toggles
this function between Auto and Manual for each offset.
*RST: Dependent upon Offset selected, please refer to the User’s
Guide for more details.
Range: 5 ms to 4 ks
Example:
SEM:OFFS:LIST:SWE:AUTO 40 ms
Key access: >More >Spectrum Emission Mask >Offset/Limits >Sweep Time