Technical data

110 N9320A Programmer’s Guide
5 Command Reference
<freq>,...[<freq>] (up to five values)
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth used in measuring the
offset pair. When set to Auto, the Res BW from the default
tables in User’s Guide are used. When set to Man, the range
of settings is the range of available Res BWs of the analyzer,
except the maximum is further limited to not exceed (Stop
Freq - Start Freq).
*RST: Dependent upon Offset label selected.
Example: SEM:OFFS:LIST:BAND 40 kHz or SEM:OFFS:LIST:BWID
40 kHz
Key access:
> More > Spectrum Emission Mask > Ref Channel > Res BW
Set the Start Freq
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STARt
<freq>,...[<freq>] (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:FREQuency:STARt?
[:SENSe]:SEMask:OFFSet:LIST:STATe
OFF|ON|0|1,...OFF|ON|0|1 (up to five values)
[:SENSe]:SEMask:OFFSet:LIST:STATe?
Specifies the inner limit (frequency closest to the carrier) for
both segments of the specified offset pair. When Start Freq
(Off) is selected, the offset pair is not measured.
Offsets that are turned off, for the currently selected offset
will return a - 999.0 when queried.
*RST: Dependent upon Offset selected, please refer to the User’s
Guide for more details.
Range: 0 Hz to the Stop Freq (for that offset) minus 10 Hz
Example: SEM:OFFS:LIST:FREQ:STAR 2 MHz
SEM:OFFS:LIST:STAT 1
Key access:
> More > Spectrum Emission Mask > Offset/Limits > Start Freq
Meas
Meas