Technical data
Command Reference 5
N9320A Programmer’s Guide 109
Set the Channel Res BW
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO?
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]
<freq>
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth used in measuring and
the power in the main channel.
*RST: 30 kHz
Range: 10% to 100% of the setting of Chan Span
Example: SEM:BAND 4 MHz
SEM:BWID 4 MHz
Key access:
> More > Spectrum Emission Mask > Ref Channel > Res BW
Set Chan Span
[:SENSe]:SEMask:FREQuency:SPAN <freq>
[:SENSe]:SEMask:FREQuency:SPAN?
Specifies the span used in measuring the power in the main
channel.
*RST: 5 MHz
Example: SEM:FREQ:SPAN 4 MHz
Key access:
> More > Spectrum Emission Mask > Chan Span
Set the Offset Res BW
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO OFF|ON|0|1,...OFF|ON|0|1
(up to five values)
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolu-
tion]:AUTO?
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]
Meas
Meas