Technical data
108 N9320A Programmer’s Guide
5 Command Reference
[:SENSe]:SEMask Subsection
Set the Average
[:SENSe]:SEMask:AVERage:COUNt <integer>
[:SENSe]:SEMask:AVERage:COUNt?
[:SENSe]:SEMask:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep.
*RST: 10/Off
Range: 1 to 1000
Example: OBW:SEM COUN 20
OBW:SEM OFF
Key access:
> More > Spectrum Emission Mask > Avg Number On Off
Set the Chan Integ BW
[:SENSe]:SEMask:BANDwidth|BWIDth:INTegration
<freq>
[:SENSe]:SEMask:BANDwidth|BWIDth:INTegration?
Specifies the integration bandwidth used in calculating the
power in the main channel.
*RST: 3.84 MHz
Range: 10% to 100% of the setting of Chan Span
Example: SEM:BAND:INT 4 MHz
SEM:BWID:INT 4 MHz
Key access:
>More>Spectrum Emission Mask>Ref Channel>Chan Integ BW
Meas
Meas