Technical data
Command Reference 5
N9320A Programmer’s Guide 107
level. For example, as the reference level changes, the input
attenuator changes to ensure that a signal at the reference
level does not exceed the Max Mixer Lvl setting.
When the TOI measurement is on, this key controls the
maximum mixer level, independent of the previous setting of
Max Mixer Lvl located under Amplitude. When the TOI
measurement is Off, the previous maximum Max Mixer Lvl is
restored. Setting Max Mixer Lvl to Auto sets the maximum
mixer level to –30 dBm.
*RST: –30 dBm
Range: –100 dBm to 0 dBm
Example: OBW:TOI COUN 20
OBW:TOI OFF
Key access:
> More > TOI > Max Mixer Lvl
Set TOI Span
[:SENSe]:TOI:FREQuency:SPAN <freq>
[:SENSe]:TOI:FREQuency:SPAN?
Specify the frequency span in which intermodulation
products are measured. If you modify the value of Span in
the base instrument Span menu, the value in the Meas menu
will be updated to reflect the new value and the
measurement will restart if it is running.
*RST: 15 MHz
Example: TOI:FREQ:SPAN 20 MHz
Key access:
> More > TOI > TOI Span
Meas
Meas