Set the Average Mode
[:SENSe]:TOI:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:TOI:AVERage:TCONrol?
Press Avg Mode to select the type of termination control used
for the averaging function to either Exp or Repeat. This
determines the averaging action after the specified number
of measurements (average count) is reached.
EXP (Exponential Averaging mode) — When you set Avg
Mode to Exp, each successive data acquisition after the
average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging
weights new data more than old data, which facilitates
tracking of slow- changing signals. The average will be
displayed at the end of each sweep.
Repeat When you set Avg Mode to Repeat, after reaching
the average count, all previous result data is cleared and
the average count is set back to 1.
*RST: EXPonential
Example: TOI:AVG:TCON EXP
Key access:
> More > TOI > Avg Mode
Set the Max Mixer Lvl
[:SENSe]:TOI:FREQuency:MIXer:RANGe[:UPPer] <num-
ber>
[:SENSe]:TOI:FREQuency:MIXer:RANGe[:UPPer]?
[:SENSe]:TOI:FREQuency:MIXer:RANGe:AUTO
OFF|ON|0|1
[:SENSe]:TOI:FREQuency:MIXer:RANGe:AUTO?
Enables you to set the relationship between the highest
signal that can be displayed (the reference level) and the
input attenuation. The relationship applies whenever the
Attenuation is set to Auto. The relationship is that the
attenuation is given by reference level minus the max mixer
Meas