Technical data

Command Reference 5
N9320A Programmer’s Guide 105
Set the X dB value
[:SENSe]:OBWidth:XDB <dB value>
[:SENSe]:OBWidth:XDB?
Enables you to specify the power level used to determine the
emission bandwidth as the number of dB down from the
highest signal point (P1), within the occupied bandwidth
span. This function is an independent calculation from the
OBW calculation. The x dB Bandwidth result is also called
the emissions bandwidth, or EBW.
*RST: –26 dB
Range: –100.0 dB through –0.1 dB
Example: OBW:XDB —50 dB
Key access: > Occupied BW > Occ BW% Pwr
[:SENSe]:TOI Subsection
Set the Average
[:SENSe]:TOI:AVERage:COUNt <integer>
[:SENSe]:TOI:AVERage:COUNt?
[:SENSe]:TOI:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOI:AVERage[:STATe]?
To specify the number of measurement averages used when
calculating the measurement result set Avg Number to ON. The
average will be displayed at the end of each sweep. Setting
Avg Number to OFF disables measurement averaging.
*RST: 10/Off
Range: 1 to 1000
Example: OBW:TOI COUN 20
OBW:TOI OFF
Key access:
> Occupied BW > Avg Number On Off
BW/
Avg
Meas